g/kg Tw= 15.6 'C ppm= 11980 pw= 12.00 hPa pws= 29.91 hPa
h= 43.2 kJ/kg
>purge
Purge started, press any key to abort.
>send
S 134 RH= 40.2 %RH T= 24.1 'C Td= 9.8 'C Tdf= 9.8 'C a=
8.8 g/m3 x= 7.5
g/kg Tw= 15.7 'C ppm= 12084 pw= 12.10 hPa pws= 30.11 hPa
h= 43.5 kJ/kg
>
Where the state of the probe is indicated by the following letters and
values:
N...xxx = Normal operation where xxx = Probe heat power
H...xxx = Purge where xxx = Sensor temperature (ºC)
S...xxx = Sensor cooling
after purge
where xxx = Sensor temperature (ºC)
A...xxx = AutoCal where xxx = Sensor temperature (ºC)
w...xxx = Sensor warming where xxx = Sensor heat power
See section General Settings on page 98 for a description of the other
serial commands.
General Settings
Changing the Quantities and Units
To change quantities and units use serial commands or the optional
display/keypad. See Table 3 on page 19 for available quantities and
Table 4 on page 20 for optional quantities.
NOTE
Only the quantities selected when ordering the device can be selected as
an output quantity.
Using Display/Keypad
Use display/keypad to select the display output quantities.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Press the ► arrow button to select Display.
3. Press ► arrow button to select Quantities.
4. Select the quantity by using the ▲▼ arrow buttons. Confirm the
selection by pressing SELECT. You can select 1 ... 3 display
quantities at a time.