Vaisala DRYCAP DMT340 SERIES Guia do Utilizador Página 98

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 169
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 97
USER'S GUIDE____________________________________________________________________
98 __________________________________________________________________ M210704EN-E
g/kg Tw= 15.6 'C ppm= 11980 pw= 12.00 hPa pws= 29.91 hPa
h= 43.2 kJ/kg
>purge
Purge started, press any key to abort.
>send
S 134 RH= 40.2 %RH T= 24.1 'C Td= 9.8 'C Tdf= 9.8 'C a=
8.8 g/m3 x= 7.5
g/kg Tw= 15.7 'C ppm= 12084 pw= 12.10 hPa pws= 30.11 hPa
h= 43.5 kJ/kg
>
Where the state of the probe is indicated by the following letters and
values:
N...xxx = Normal operation where xxx = Probe heat power
H...xxx = Purge where xxx = Sensor temperature (ºC)
S...xxx = Sensor cooling
after purge
where xxx = Sensor temperature (ºC)
A...xxx = AutoCal where xxx = Sensor temperature (ºC)
w...xxx = Sensor warming where xxx = Sensor heat power
See section General Settings on page 98 for a description of the other
serial commands.
General Settings
Changing the Quantities and Units
To change quantities and units use serial commands or the optional
display/keypad. See Table 3 on page 19 for available quantities and
Table 4 on page 20 for optional quantities.
NOTE
Only the quantities selected when ordering the device can be selected as
an output quantity.
Using Display/Keypad
Use display/keypad to select the display output quantities.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Press the arrow button to select Display.
3. Press arrow button to select Quantities.
4. Select the quantity by using the ▲▼ arrow buttons. Confirm the
selection by pressing SELECT. You can select 1 ... 3 display
quantities at a time.
Vista de página 97
1 2 ... 93 94 95 96 97 98 99 100 101 102 103 ... 168 169

Comentários a estes Manuais

Sem comentários