Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 131
If the adjustment in the AutoCal reaches a preset maximum value or if
the AutoCal correction fails, for example, because of unstable conditions,
a new AutoCal will take place later (if the automatic AutoCal is turned
on).
Automatic AutoCal
As a default, the automatic AutoCal in DMT340 is turned on. In this
mode, the calibration takes place automatically if the dewpoint or
temperature changes significantly, typically more than 10 ºC. However, if
there are no changes in the conditions, the AutoCal will take place
repeatedly after one hour from the last AutoCal.
Manual AutoCal
To check that AutoCal has taken place for obtaining the most accurate
measurement in a very dry environment, you can perform Autocal
manually before measuring as follows:
1. Open the MAIN MENU by pressing any of the arrow buttons.
2. Select Measuring, press the ► arrow button.
3. Select Manual AutoCal, press the ► arrow button.
4. Press START to start AutoCal. If the AutoCal conditions are not
fulfilled, a note appears on the display informing that the
calibration cannot be made.
0706-014
Figure 70 Following AutoCal on the Display
5. Press EXIT to return to the basic display.
You can also perfom Autocal manually by using the pressure set and
check buttons on the motherboard. By pressing both buttons
simultaneously, Autocal is performed, see Figure 64 on page 102.
Sensor Purge
Sensor Purge is a feature available for the DMT340 M-sensor. The Purge
should be carried out to achieve the shortest response times and the best
long-term stability.