USER'S GUIDE____________________________________________________________________
52 __________________________________________________________________ M210704EN-E
Sampling Cell for DMT348
It may be necessary to use a sampling cell if the process (for example a
pipe) is too small for the DMT348 probe. Furthermore, if the process is
very hot (>80 °C) or particularly dirty, the probe is installed in a
sampling cell behind a cooling coil and/or filter. In this case, the ambient
temperature must be at least 10 °C warmer than the process dewpoint in
order to avoid condensation in the sample tubing.
Sampling Cell with Swagelok Connectors (Vaisala order code:
DMT242SC2) and Sampling Cell with Female Connectors (Vaisala order
code: DMT242SC) are available as an option.
0506-026
Figure 34 Sampling Cells DMT242SC2 and DMT242SC
Numbers refer to Figure 34 above:
1 =
Male pipe weld connector Swagelok 1/4"
2 = G1/2"
3 = G1/4"
4 = G3/8"