Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 111
Data Recording
Data recording function is always on and collects data automatically into
the memory of the device. If the optional data logger module is installed,
the transmitter uses it automatically. Recorded data does not disappear
from the memory when the power is switched off. Collected data can be
observed in a form of a graph in the graphical view of the display or it
can be listed out by using the serial line or MI70 Link program.
Selecting the Data Recording Quantities
If the device is provided with the optional display, the recorded quantities
are always those selected for the display. Up to three quantities can be
recorded at a time. See section Changing the Quantities and Units on
page 98 for instructions on how to select the display quantities with the
keypad.
DSEL
Use the serial line command DSEL [xxx] to select the quantities to be
recorded if the transmitter is not equipped with display/keypad.
You may also dump the logged data to the serial line in numeric form
with the following commands.
DSEL [xxx]
where
xxx = Data recording quantity. See Table 3 and Table 4 on page 19
for the quantities.
Enter the command without parame
ters and press ENTER to display
current recording parameters.
Example:
>dsel rh t tdf
RH T Tdf
View Recorded Data
If the device is provided with the optional display, the graphical display
shows the data of the selected quantities, one at a time. See section
Graphic History on page 72 for details about graphical display.