Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 117
8. Select the upper limit by pressing the ▲▼arrow buttons. Use the
arrow buttons to set the upper limit value. Press SET to confirm
your selection. Press OK to confirm your setting.
9. Press EXIT to return to the basic display.
AMODE/ASEL
Use the serial line to select and scale the analog output quantities.
Connect the transmitter to the PC. Open the terminal connection between
your PC and the transmitter.
1. Check the analog outputs with the AMODE command.
Example:
>amode
Ch1 output : 0...1V
Ch2 output : 0...1V
>
2. Select and scale the quantities for the analog outputs with the
command ASEL [xxx yyy zzz]. Note that the optional quantities can
be selected only if they have been selected when ordering the
device.
where
xxx
= Quantity of channel 1
y
yy
= Quantity of channel 2
zzz
= Quantity of the optional analog output channel 3
Always enter all the quantities for all outputs. For quantities and their
abbreviations see Table 3 on page 19.
Use the command ASE
L [xxx yyy] as shown in the example below when
using a device with two analog outputs.
Example:
>asel td t
Ch1 Td low : -20.00 'C ?
Ch1 Td high : 100.00 'C ?
Ch2 T low : -40.00 'C ?
Ch2 T high : 180.00 'C ?
>