
USER'S GUIDE____________________________________________________________________
124 _________________________________________________________________ M210704EN-E
Operation of the Indication LEDs
Relay is activated: LED is lit.
Relay is not activated: LED is not lit.
Setting the Relay Outputs
NOTE
When having only one relay module installed, its relays are called 'relay
1' and 'relay 2'.
When having two relay modules, the relays of the the module connected
to slot MODULE 1 (communications) are called 'relay 1' and relay 2' and
relays connected to slot MODULE 2 are called 'relay 3' and 'relay 4'
0706-013
Figure 69 Relay Availability
Number refers to Figure 69 above:
1 = Lists enabled relays. Activation state shown in black. Disabled
relays are not shown.
Use the display/keypad to set the relay outputs.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select Interfaces, confirm by pressing the ► arrow button.
3. Select Relay outputs, confirm by pressing the ► arrow button.
4. Select Relay 1/2/3/4, confirm by pressing the ► arrow button.
5. Select the Quantity, confirm by pressing Change. Select the
Quantity by using the arrow buttons. Confirm your selection by
pressing Select. (Press Fault Status when the relay follows the
transmitter error.) Press Change to to set the value.
6. Select Act. above / Act. below. Press SET to confirm your
selection. If asked, select MODIFY if you want to set the setpoint
Comentários a estes Manuais