Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 101
where
x
= M or N
where
M = Metric units
N = Non-metric units
NOTE
This command changes both the serial output and display units to either
metric or non-metric units. When you want to output both metric and
non-metric units simultaneously on the display, select the display units
by using the display/keypad.
Use UNIT H
2
O [ppmv/ppmw] to change H
2
O units.
Use the MOL/MOLI command to view/set mole weight parameter that
will be used in calculating H
2
O ppm
w.
Pressure Compensation Setting
The pressure has an effect on humidity calculations and accuracy.
Therefore, accurate calculations can be achieved only when the process
pressure is taken into consideration.
Note that conversions from mmHg and inHg are defined at 0°C and for
mmH
2
O and inH
2
O at 4°C.
NOTE
Pressure compensation is intended to be used in normal air only. When
measuring in other gases, please contact Vaisala for further information.
NOTE
Fixed pressure compensation value of 1013.25 hPa is used when in
adjustment mode.
Using Display/Keypad
Use display/keypad to set the pressure compensation. To select the
pressure unit using display/keypad see section Changing the Quantities
and Units on page 98.
1. Press any of the arrow buttons to open the MAIN MENU.