Vaisala DRYCAP DMT340 SERIES Guia do Utilizador Página 45

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 169
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 44
Chapter 3 _______________________________________________________________ Installation
VAISALA _______________________________________________________________________ 45
Tightening the Clasp Nut
1. Adjust the probe to a suitable depth according to the type of
installation.
2. Tighten the clasp nut first manually.
3. Mark the fitting screw and the clasp nut.
4. Tighten the nut a further 50 … 60º (ca. 1/6 turn) with a wrench. If
you have suitable torque spanner, tighten the nut to max 45 ± 5 Nm
(33 ± 4 ft-lbs).
0505-276
Figure 29 Tightening the Clasp Nut
The following numbers refer to Figure 29 above:
1 = Probe
2 = Clasp nut
3 = Fitting screw
4 = Pen
NOTE
Take care not to over tighten the clasp nut to avoid difficulties when
opening it.
CAUTION
Take care not to damage the probe body. A damaged body makes the
probe less tight and may prevent it from going through the clasp nut.
Vista de página 44
1 2 ... 40 41 42 43 44 45 46 47 48 49 50 ... 168 169

Comentários a estes Manuais

Sem comentários