Vaisala DRYCAP DMT340 SERIES Guia do Utilizador Página 122

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 169
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 121
USER'S GUIDE____________________________________________________________________
122 _________________________________________________________________ M210704EN-E
Relay Indicating Transmitter Error Status
You can set a relay to follow the operation status of the device. By
selecting FAULT/ONLINE STATUS for output quantity a relay changes
state on the basis of the operation status as follows:
FAULT STATUS
Normal operation: relay active (C and NO outputs are closed)
Not measuring state (error state or power off): relay released (C and NC
outputs are closed)
ONLINE STATUS
Live measurement (data available): relay active (C and NO outputs are
closed)
No live data (for example: error state, chemical purge or adjustment
mode): relay released (C and NC outputs are closed)
See Figure 68 on page 123 for illustrative examples of the
FAULT/ONLINE STATUSrelay output modes.