Vaisala BAROCAP PTB330 Guia do Utilizador Página 77

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 138
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 76
Chapter 4 ________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 75
HQFE [-30...+30 m] Shows or sets the altitude for the QFE
corrected pressure.
HQNH [-30...3000 m] Shows or sets the altitude for the QNH
corrected pressure.
PSTAB Shows or sets the pressure stability
limits.
AVRG Sets the barometer measurement
averaging time (in seconds).
Formatting commands
FORM Sets the custom output for the SEND
command and for RUN mode.
TIME Shows or changes the current time.
DATE Shows or changes the current date set.
UNIT Shows or sets unit for a quantity. If
several quantities offer the same unit,
these will also be changed to the given
unit.
Data recording commands
DSEL Selects the quantities that will be
displayed on the graphical user interface.
The quantities on the display will also be
logged.
DELETE Erases the log memory.
UNDELETE Restores the erased log memory.
DIR Lists the available logs in the logging
memory.
PLAY Shows the trend, min and max values of
the given log.
Calibration and adjustment commands
CDATE Shows or sets the calibration date.
LCP1/
LCP2/
LCP3
Performs a linear correction for the
barometer module/module.
Vista de página 76