Vaisala BAROCAP PTB330 Guia do Utilizador Página 111

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 138
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 110
Chapter 6 __________________________________________________ Calibration and Adjustment
VAISALA______________________________________________________________________ 109
0706-072
Figure 46 Adjustment Menu
Pressure Adjustment
In PTB330, the raw measured pressure value of each barometer module
is first corrected using the corresponding multipoint adjustment values.
The additional linear adjustment (if enabled) is applied to the corrected
values. This allows the user to fine-tune the pressure display using the
linear adjustment without performing a complete readjustment for the
barometer module, see Figure 47 on page 109.
0708-053
Figure 47 PTB330 Adjustments
For correct adjustment, the existing adjustment must be disabled before
measuring the new adjustment data. If you are going to do only linear
adjustment (one to two pressure points), leave the multipoint
adjustments active.
If you want to perform a complete readjustment of your PTB330,
disable both the multipoint and the linear adjustments, and calibrate the
instrument at 2…8 pressure points. In such a case, the new adjustment
values are entered to the multipoint adjustment function and the linear
adjustment is not used (that is, it is left disabled).
Vista de página 110
1 2 ... 106 107 108 109 110 111 112 113 114 115 116 ... 137 138

Comentários a estes Manuais

Sem comentários