Vaisala BAROCAP PTB330 Guia do Utilizador Página 120

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 138
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 119
User's Guide ______________________________________________________________________
118 _________________________________________________________________ M210855EN-C
*Defined as ±2 standard deviation limits of endpoint non-linearity,
hysteresis or repeatability error.
**Defined as ±2 standard deviation limits of inaccuracy of the working
standard including traceability to NIST.***Defined as the root sum of
the squares (RSS) of endpoint non-linearity, hysteresis error,
repeatability error and calibration uncertainty at room temperature
****Defined as ±2 standard deviation limits of temperature dependence
over the operating temperature range.
Operating Environment
Table 18 Temperature Dependence****
Class B
500 ... 1100 hPa ±0.01 hPa
50 ... 1100 hPa ±0.03 hPa
Table 19 Total Accuracy at -40 ... +60°C (-40 ... +140 °F)
Class A Class B
500 ... 1100 hPa ±0.15 hPa ±0.25 hPa
50 ... 1100 hPa ±0.45 hPa
Table 20 Long Term Stability
500 ... 1100 hPa ±0.10 hPa/year
50 ... 1100 hPa ±0.20 hPa/year
Table 21 Operating Environment
Pressure range 500 ... 1100 hPa, 50 ... 1100 hPa
Temperature range
operating -40 ... +60°C (-40 ... +140 °F)
with local display 0 ... +60°C (-32 ... +140 °F)
Maximum pressure limit 5000 hPa abs.
Electromagnetic compatibility EMC standard EN61326-1:1997+
Am1:1998 + Am2:2001: Industrial
Environment
Vista de página 119
1 2 ... 115 116 117 118 119 120 121 122 123 124 125 ... 137 138

Comentários a estes Manuais

Sem comentários