Vaisala BAROCAP PTB330 Guia do Utilizador Página 37

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 138
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 36
Chapter 3 _______________________________________________________________ Installation
VAISALA_______________________________________________________________________ 35
D-9 Connector
0712-017
Figure 19 Wiring of Optional D-9 Connector
Table 3 Pin Assignments to RS-232/485 Serial Output
Pin Wire
Color
Serial Signal Analog Signal
+RS232
RS-232 (EIA-232) RS-485 (EIA-485)
1Red
2 White TX TX
3 Black RX RX
4 (optional) Yellow External power control External power control External power control
5 Brown Ground Ground
6 Green LO Aout
7 Blue Ground for supply voltage Ground for supply voltage GND supply
8Grey HI AGND
9 Orange Supply voltage (10...30
VDC)
Supply voltage (10...30
VDC)
Supply
NOTE
The D-9 connector cannot be used with relay modules or power supply
module that have AC (mains) power connection.
NOTE
The D-9 connector is not IP65 protection classified.
Vista de página 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 137 138

Comentários a estes Manuais

Sem comentários