Vaisala WMS302 Guia do Utilizador Página 172

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 171
User's Guide ______________________________________________________________________
170 __________________________________________________________________M210630EN-B
Anemometer bearings:
1. Loosen the hubnut (14) with fingers or a 10 mm tool and remove
the cup wheel (13).
2. Remove the ball bearing assembly (12) by unscrewing it
counterclockwise (with a 10 mm tool).
3. Insert a new bearing assembly (12). Tighten gently.
4. Fasten the cup wheel to the sensor. Tighten gently.
Vane bearings:
1. Proceed as described in steps 1 and 2 above.
2. Open the lock screw (11) of the tail assembly (10) and remove the
screw.
3. Remove the Seeger-ring (9) (with narrow point pliers).
4. Remove the bearing assembly (8).
5. Replace the bearings inside the housing with new ones.
6. Assemble the sensor in the reverse work order.
Note that part number 6 in Figure 143 on page 171 consists of two parts.
The lower portion has a notch that is aligned at the factory in relation to
the output of the potentiometer to point north when the sensor is
mounted.