Accredited Humidity & Temperature Calibration
11604121 – Rev. B
ACCREDITED HUMIDITY CALIBRATION
The relative humidity calibration is performed at room
temperature using a Thunder Scientific 2500 Two-
Pressure Generator. The calibration uncertainty
typically ranges from ± 0.60 to ±1.1 %RH. A one-
point temperature calibration is performed at ambient.
The calibration procedure used for calibrations is
11603108-CSL Humidity Calibration.
Thunder Scientific 2500 Two-Pressure Generator
saturates a continuous stream of gas with water vapor
at a known pressure and temperature, and passes it
through an expansion valve where it expands to a
lower pressure.
ACCREDITED TEMPERATURE CALIBRATION
The temperature calibration is made using Hart
Scientific liquid bath with a Black Stack reference.
The calibration uncertainty is typically ±0.02 °C. The
calibration procedure used for calibration is 11603135-
CSL Temperature Calibration.
Hart Scientific liquid baths are used as the calibration
method. A Hart Scientific Black Stack/PRT
combination is used as the reference standard.
The
Black Stack uses a high accuracy display module,
model 1560 as the readout indicator. The PRT is
immersed in the liquid of the bath.
CALIBRATION DETAILS
The instrument will be adjusted to comply with its
specifications (if needed) unless no adjustment is
requested. Out of tolerance conditions are determined
by the product specification only. The calibration
Calibration Laboratory Certificate #: 2083.01
ISO/IEC 17025:2005
CALIBRATION DETAILS CONTINUED
uncertainty is not tied in with the instrument's
accuracy. All results are reported in a calibration
certificate, which is written according to the ISO/IEC
17025:2005 standard.
ACCREDITATION
Accreditation is a third party formal recognition that
a laboratory is competent to carry out specific
calibrations.
The American Association for Laboratory
Accreditation (A2LA) has assessed Vaisala’s
Calibration Standards Laboratory (CSL) for humidity
and temperature calibrations. The
accreditation
standard used is ISO/IEC 17025:2005. A2LA
accreditation is recognized by all the major
international accreditation organizations.
International
Level
National
Level (USA)
Secondary
Level
Customer
Level
Traceability chain for humidity calibration
Traceability chain for temperature calibration
SI – International S
y
stem of Units
National Institute of Standards and
Technolo
gy
(
NIST
)
SI – International S
y
stem of Units
National Institute of Standards and
Technolo
gy
(
NIST
)
Hart Scientific
CSL Primary Std.: Hart Scientific
Su
p
er-Thermometer/SPRT
Calibration for Customer
Workin
g
Standards
Calibration for Customer
Workin
g
Standards
CSL Standard: Hart Scientific
Black Stack/PRT
CSL Standard: Thunder Scientific
2500 Generator
CSL Primary Standard: MBW
Chilled Mirror H
yg
rometer
Accredited Humidity & Temperature Calibration
11604121 – Rev. B
Ordering a calibration that complies with the requirements of ISO/IEC 17025:2005
Calibration for Vaisala humidity and temperature transmitters and probes can easily be ordered using this
form. Please read the form and notes carefully and complete the form accurately to ensure correct
calibration service. The form must be completed in full before calibration services can start. Include the
form with your instrument. For additional information please contact Vaisala at 1-800-408-9454.
Instrument Details
Customer Name:
Model No(s):
Serial No(s):
Calibration Details Price
Humidity Calibration
Performed at ambient temperature. Includes 1-point ambient temperature calibration
.
Extra points are $95 each (example: 7 points = $545). Please specify custom points below.
0 Points No Humidity Calibration
A
$0.00
4 Points 0, 11, 50, and 80 %RH
B
$325.00
6 Points 0, 11, 30, 50, 70, and 90 %RH
C
$450.00
8 Points 0, 11, 30, 50, 60, 70, 80, and 90 %RH
D
$600.00
10 Points 0, 11, 30, 40, 50, 60, 70, 80, 90, and 95 %RH
E
$750.00
Temperature Calibration - Specify Points Below
Test points must be between -80 °C and 180 °C and within the model’s measurement range.
0 Points
No Temperature Calibration
A
$0.00
3 Points
1_____ 2_____ 3_____ °C / °F
B
$225.00
5 Points
1_____ 2_____ 3_____ 4_____ 5_____ °C / °F
C
$350.00
7 Points
1_____ 2_____ 3_____ 4_____ 5_____ 6_____ 7_____ °C / °F
D
$450.00
9 Points 1_____ 2_____ 3_____ 4_____ 5_____ 6_____ 7_____ 8_____ 9_____ °C / °F
E
$550.00
Units of Results Reported
Humidity %RH, Temperature °C
A
$0.00
Humidity %RH, Temperature °F
B
$0.00
Adjustment
Calibration prices include adjustment (if needed). Check here if you DO NOT want your instrument adjusted.
Customer Information To Appear On Certificate
Company Name (if different than above):
Address:
Requested Due Date For Next Calibration:
Customer Asset #:
Received And Accepted At Vaisala
Name & Date:
Contract #:
NOTE: The calibration prices include a Calibration Certificate, but do not reflect repair charges.
Repair charges will utilize fixed pricing: Analog repairs: $125.00, Digital repairs: $350.00