DMT142 Miniature Dewpoint Transmitter
for OEM Applications
The Vaisala DRYCAP
®
Dewpoint Transmitter DMT142 is an ideal choice for
small compressed air dryers, plastic dryers and other OEM applications.
Features/Benefits
Miniature size dew point
transmitter for e.g. small
industrial dryer applications
Vaisala DRYCAP® technology
with auto-calibration
Calibration interval of two years
Dew point measurement range
-60 ... +60 °C (-76 ... +140 °F)
with an accuracy ±3 °C (±5.4°F)
Withstands condensation
Fast response time
Compatible with Vaisala
DRYCAP
®
Hand-Held Dewpoint
Meter DM70
NIST traceable (certificate
included)
Vaisala DRYCAP®
The Vaisala DRYCAP
®
Dewpoint
Transmitter DMT142 is a miniature
dew point measurement instrument.
The transmitter can be installed
directly into pressurized systems at
50 bar (725 psia) maximum pressure.
The long-term high performance
is achieved with Vaisala DRYCAP
®
technology.
The sensor fully withstands getting
wet, and therefore, the transmitter
performs exceptionally well in
applications that occasionally
experience process water spikes,
such as pipeline condensation during
a system failure or start-up. The
sensor is also immune to particulate
contamination, oil vapor and most
chemicals, and is insensitive to the
ow rate.
Long calibration interval
The calibration interval of the
DMT142 is two years. Additionally,
the Vaisala DRYCAP
®
Hand-Held
Dewpoint Meter DM70 can be used
to conrm the performance of the
DMT142 without disconnecting the
transmitter. For any adjustment
needs, the transmitter can be sent to
Vaisala Service.
The auto-calibration software works
on-line while the process is running.
If the measurement accuracy is not
conrmed, corrections are made
automatically.
Easy installation
The DMT142 has a variety of features
to choose from, including different
output and installation options. Due
to its small size and light weight,
the DMT142 is quickly and easily
installed in tight spaces or in small-
size pipelines.
The Vaisala DRYCAP
®
Hand-Held
Dewpoint Meter DM70 is ideal for
confirming the performance of the
DMT142 in the field.
Technical data
Ref. B210375EN-B ©Vaisala 2009
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
Measured variables
DEW POINT TEMPERATURE
Measurement range(typical)
Standard version -40 ... +60 °C (-40 ... +140 °F) T
d
Extended version -60 ... +60 °C (-76 ... +140 °F) T
d
SF
6
version -50 ... +60 °C (-58 ... +140 °F) T
d
Analog output scalings
option 1 -80 ...+20 °C (-112 ... +68 °F) T
d
option 2 -60 ... +40 °C (-76 ... +104 °F) T
d
option 3 free scaling
Accuracy ±3 °C (±5.4 F) T
d
(see graph below)
(when the dew point is below 0 °C (32 °F), the transmitter outputs
frostpoint)
Operating environment
Temperature *) -40 ... +60 °C (-40 ... +140 °F)
Relative humidity 0 ... 100 %RH
Pressure *) 0 ... 50 bar
a
(725 psia)
Sample flow rate no effect for measurement accuracy
*) For extended temperature down to -40 °C (-40 °F) or pressure up
to 50 bar
a
(725 psia) the supply voltage must be 24 ... 28 VDC.
Outputs
Analog output (scalable) 4 ... 20 mA (3-wire), 0 ... 1 V/5 V
Resolution for current output 0.002 mA
Resolution for voltage output 0.3 mV
Typical temperature dependence 0.005% of span / °C
Connector 4-pin M8 (IEC 60947-5-2)
connection cable with snap-on or thread locking available
RS232 serial line for service use with DMT142RS cable
General
Sensor Vaisala DRYCAP
®
180D
Measured gases non-corrosive gases
(SF
6
gas with special model)
Recommended calibration interval
to confirm the specified accuracy 2 years
Operating voltage with voltage output 12 ... 28 VDC
Operating voltage with current output 18 ... 28 VDC
Supply current
normal measurement 10 mA + load current
during self-diagnostics max. 220 mA pulsed
Load for current output max. 500 ohm
Load for voltage output min. 10 kohm
Housing material stainless steel body (AISI316L)
plastic cap(ABS/PC)
Sensor protection stainless steel sintered filter
Mechanical connection G1/2” ISO 228-1 thread
with bonded seal ring
(U-seal)
Housing classification IP64 (NEMA 3S)
Storage temperature range -40 ... +80 °C (-40 ... +176 °F)
Weight 118 g (4.16 oz)
Complies with EMC standard EN61326-1, Electrical equipment
for measurement, control and laboratory use - EMC requirements;
Industrial environment.
DEW POINT ACCURACY VS. MEASUREMENT CONDITIONS
Dew point accuracy range for the extended version. Standard
version with dotted line.
Response time 63% [90%] at +20 °C gas temperature and 1 bar
pressure
-60 -> -20 °C T
d
(-76 -> -4 °F T
d
) 5 s [10 s]
-20 -> -60 °C T
d
(-4 -> -76 °F T
d
) 45 s [10 min]
PPM VOLUME CONCENTRATION
Measurement range 70 ... 200 000 ppm
Accuracy at +20 °C (+68 °F), 1013 mbar 7.3 ppm + 9.2% of reading
Accessories
Output cable M8, snap-on connector, 2 meters 211598
Output cable M8, thread connector, 3 meters HMP50Z300, several
lengths available
Connection cable for DM70 211917ZZ
Service cable for serial line DMT142RS
Sampling cells
basic sampling cell DMT242SC
with Swagelok 1/4" male connectors DMT242SC2
with quick connector and leak screw DSC74
two-pressure sampling cell DSC74B
cooling/venting coil DMCOIL
See DM70 / Portable Sampling Systems and Sampling Cells for
further information about sampling cells available