76 77
DMT242 Dewpoint Transmitter for
OEM Applications
Due to its wide measurement range and high long-term stability, the DMT242 is an
ideal choice for low dewpoint industrial applications such as compressed air dryers,
plastic dryers and other OEM applications.
Features/Benefits
Ideal choice for industrial dryer
applications
Incorporates advanced Vaisala
DRYCAP® Sensor and enhanced
auto-calibration software
Long-term stability in low
dewpoints
Fast response time
Two sensor options cover
dewpoint measurement range
from -60 … +60 °C
(-76 … +140 °F) with an
accuracy of ±2 °C (±3.6 °F)
Withstands condensation
NIST traceable (certificate
included)
Compatible with Vaisala
DRYCAP® Hand-Held Dewpoint
Meter DM70
Vaisala DRYCAP
®
The Vaisala DRYCAP
®
Dewpoint
Transmitter DMT242 provides
reliable and stable measurements
for industrial dryer applications. It is
designed for extreme conditions.
DMT242 incorporates the Vaisala
DRYCAP
®
thin lm polymer sensor
and auto-calibration software.
The standard sensor choice for
dry gases and desiccant dryers is
DRYCAP
®
180M and for more humid
applications such as regeration
dryers, a DRYCAP
®
180S sensor.
Both the sensors are immune to
particulate contamination, water
condensation, oil vapor and most
chemicals. Because the sensor
withstands condensation, its
performance is unmatched for
low dewpoint applications that
experience process water spikes,
such as pipeline condensation during
a system failure or start-up.
The auto-calibration software works
on-line while the process is running.
If the measurement accuracy is not
conrmed, corrections are made
automatically. The DMT242 adjusts
the measurement, corrects dry-end
drifts and continues to function.
Calibration occurs quickly, and
with corrections so minor, it will go
unnoticed.
Compact, rugged and
intelligent
Due to its compact size, DMT242 is
quickly and easily installed in tight
spaces.
Users can perform a eld-check by
using the Vaisala DRYCAP
®
Hand-
Held Dewpoint Meter DM70. The
transmitter can be sent to Vaisala
Service for NIST traceable calibration.
The recommended calibration
interval is every two years.
The Vaisala DRYCAP
®
Hand-Held Dewpoint
Meter DM70 is ideal for confirming the
performance of the DMT242 in the field.
Technical data
Ref. B210765EN-B ©Vaisala 2009
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
Dewpoint temperature (°C)
0
20 40 60
0
40
-60
-80
Process temperature (°C)
10
20
30
50
60
70
-10
-20
-30
-50
-40
-70
3010
50
Accuracy < ± 2°C Td
Dewpoint measurement range lower
limit
Water vapor saturation line
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
Dewpoint temperature
Measurement range (typical) -60 ... +60 °C (-76 ... +140 °F)
Analog output scalings
Option A -80 ...+20 °C (-112 ... +68 °F) T
d
Option B -60 ... +60 °C (-76 ... +140 °F) T
d
Option X free scaling
(when the dewpoint is below 0 °C (32 °F),
the transmitter outputs frostpoint)
Accuracy with DRYCAP
®
180M ±2 °C (±3.6 °F)
(see graph below)
Operating environment
Temperature 0 ... +60 °C (32 ... +140 °F)
higher temperature peaks Short-term OK
Relative humidity 0 ... 100 %RH
Pressure 0 ... 20 bara (0 ... 290 psia)
Sample flow rate no effect
Output
Analog output 4 ...20 mA
Resolution for analog output ±0.002 mA
Typical temperature dependence 0.0008 mA/ °C
Serial line for service use RS232
General
Sensor DRYCAP
®
180M
Optimal sensor for refrigeration dryers DRYCAP
®
180S
Operating voltage 18 - 35 VDC, 20 - 28 VAC
Power consumption at 24 VDC max. 220 mA
External load for analog output max. 500 
Optional connection cable with
DMT242 connector 2 m or 10 m
Connector for supply voltage and
signal output
max. wire size 0.75 mm
2
max. cable diameter 6.5 mm /PG7
Service cable for serial interface
RS232 product code DMT242RS
Probe material (wetted parts) stainless steel
(AISI 316L)
Sensor protection stainless steel
sintered filter (HM47280)
Mechanical connection G½” ISO228-1 thread with
bonded seal ring (U-seal)
Eletronics housing material plastic (ABS/PC)
Housing classification IP65 (NEMA4)
Storage temperature range -40 ... +70 °C, (-40 ... +158 °F)
Complies with the EMC standard EN61326-1, Electrical equipment
for measurement, control and laboratory use - EMC requirements;
Industrial environment.
DEWPOINT ACCURACY VS. MEASUREMENT CONDITIONS
Response time 63 % [90 %] at +20 °C gas temperature
Flow rate >1 l/min and 1 bar pressure
-60 -> -20 °C Td (-76 -> -4 °F Td) 5 s [10 s]
-20 -> -60 °C Td (-4 -> -76 °F Td) 45 s [10 min]
Accuracy with DRYCAP
®
180S ±2 °C (±3.6 °F)
(see graph below)
Vista de página 45