CHAPTER 6
CALIBRATION AND ADJUSTMENT
This chapter provides information that is needed in calibration and
adjustment of the product.
NOTE
Calibration means comparing the instrument to a known reference,
either against a second instrument or a known reference environment.
Correcting the reading of the instrument so that is measures accurately is
referred to as adjustment.
NOTE
It is important to clean the sensor before calibration as the oily sensor can
contaminate the salt bath and change the reference condition. For
instructions, see section Cleaning on page 50.
The MMT162 is fully calibrated and adjusted as shipped from factory.
The calibration interval depends on the application. You can perform a
calibration to check that the transmitter is within its accuracy
specification.
Performing an accurate calibration and adjustment takes some time and
preparation. Instead of doing it yourself, you can also have a Vaisala
service center calibrate and adjust your transmitter. For contact
information, see section Product Returns on page 53.