CHAPTER 4
OPERATION
This chapter contains information that is needed to operate this product.
Transmitter Start-Up
When the transmitter is connected to power supply, there is a delay of
five seconds as the transmitter starts up. After the start-up is complete,
the measurement data is available from the analog outputs or serial line.
If serial line output is used, note that the transmitter will behave
according to the configured serial mode:
- In STOP mode, the transmitter outputs the transmitter model and
software version. This is the default mode.
- In RUN mode, a measurement output starts immediately.
- In POLL mode, the transmitter does not output anything after power-
up.
- In MODBUS mode, the transmitter does not output anything after
power-up.
For a description of the modes, see section Set Serial Interface Mode on
page 33.
Analog Output Operation Modes
The MMT162 has two analog output channels. These channels can
function in two different modes, which can be set individually for both
channels:
1. Normal operation mode
2. Limit alarm mode
When the transmitter is ordered from Vaisala, the modes are configured
as specified in the order form. You can change the operation settings
after delivery by using serial commands and the USB cable accessory.