Vaisala HUMICAP MMT162 Guia do Utilizador Página 37

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 36
Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 35
Example:
amode 1 2
Ch1 output : 0 ... 20 mA
Ch2 output : 4 ... 20 mA
Calibrate Analog Output
Use the ACAL command to adjust analog output channels.
ACAL [channel]<cr>
where
channel
=
Analog output channel to be adjusted, 1 or 2.
After entering the ACAL command, use a calibrated current/voltage
meter to measure the output, and enter the values.
Example:
acal 1
Ch1 U1 ( V ) ? 1.001
Ch1 U2 ( V ) ? 9.011
acal 2
Ch2 U1 ( V ) ? 0.0988
Ch2 U2 ( V ) ? 0.8997
Set Malfunction Alarm Output for Analog
Channels
Use the AERR command to set the analog output levels that are output in
case of a device malfunction. The default output level is 0.
AERR<cr>
Example:
aerr
Ch1 error out : 10.000V ? 0
Ch2 error out : 1.000V ? 0
NOTE
The error output value must be within a valid range of the output mode.
NOTE
The error output value is displayed
only when there are minor electrical
faults such as a humidity sensor damage. When there is a severe device
malfunction, the error output value is not necessarily shown.
Vista de página 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 74 75

Comentários a estes Manuais

Sem comentários