HMP260 SERIESTRANSMITTERSOperating ManualHMP260-U017en-1.55 January 1996 Vaisala 1996

Página 2 - Back of cover page

HMP260 SERIESOperating Manual HMP260-U017en-1.54 1997-11-253. INSTALLATION3.1. Selecting the place of installationThe transmitters should be installed

Página 3 - Contents

HMP260 SERIESHMP260-U017en-1.5 Appendix 3: RS 485/422 serial port module5 SPECIFICATIONSConnections Berg socketson the main board screw terminals 0.5

Página 4

HMP260 SERIESHMP260-U017en-1.5 Appendix 4: Digital current loop moduleINSTALLING AND USING THE DIGITAL CURRENT LOOP MODULEINSTALLING AND USING THE DIG

Página 5

HMP260 SERIESAppendix 4: Digital current loop module HMP260-U017en-1.51 INSTALLATIONSwitch the transmitter off.Resistors R2, R3 and R4 between connect

Página 6

HMP260 SERIESHMP260-U017en-1.5 Appendix 4: Digital current loop module2 OPERATIONThe HMP260 transmitters can either be given an address or operated wi

Página 7 - 1. PRODUCT DESCRIPTION

HMP260 SERIESAppendix 4: Digital current loop module HMP260-U017en-1.5The digital current loop module has separate lines for transmitting and receivin

Página 8 - 2 1997-11-25

HMP260 SERIESHMP260-U017en-1.5 Appendix 4: Digital current loop module3 NETWORK CONFIGURATIONSingle loop operationBidirectional data on one pair and g

Página 9 - 1997-11-25 3

HMP260 SERIESAppendix 4: Digital current loop module HMP260-U017en-1.5• Set the transmitter in POLL mode:>smode pollSerial mode : POLLNOTE 1The S

Página 10 - 3. INSTALLATION

HMP260 SERIESHMP260-U017en-1.5 Appendix 4: Digital current loop moduleDual loop wiring• Set the address of the transmitter, it can be any number betwe

Página 11 - 1997-11-25 5

HMP260 SERIESAppendix 4: Digital current loop module HMP260-U017en-1.54 CHECKING THE SERIAL PORT NETWORK OPERATIONNormally measurement readings are as

Página 12 - 6 1997-11-25

HMP260 SERIESHMP260-U017en-1.5 Appendix 4: Digital current loop modulePOLL modeIf a transmitter has been set to POLL mode, it will respond only to com

Página 13 - 1997-11-25 7

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 53.2. MountingHMP263/264/265(EEx ia) IICSAFE AREA HAZARDOUS AREA EEx ia IIC T3/T4HMP263 s

Página 14 - 8 1997-11-25

HMP260 SERIESAppendix 4: Digital current loop module HMP260-U017en-1.55 SPECIFICATIONSGalvanic isolation 1500 VAC/DC max. (1 min)Loop supply voltage 4

Página 15 - 1997-11-25 9

HMP260 SERIESHMP260-U017en-1.5 Appendix 5: Error messagesERROR MESSAGESThe HMP260 transmitters go through a self-diagnostics procedure when the power

Página 16 - 10 1997-11-25

HMP260 SERIESAppendix 5: Error messages HMP260-U017en-1.5Errors during operationThere are two types of errors that may occur during the operation of t

Página 17 - SAFE AREA

HMP260 SERIESHMP260-U017en-1.5 Appendix 5: Error messagesThe LEDs indicate these errors as follows:¡¡¤¤ no frequency at all¡¡¤¡ a frequency missing fr

Página 18

HMP260 SERIESHMP260-U017en-1.5 Appendix 6: Calculation formulas1 CALCULATIONThe HMP260 series transmitters measure relative humidity and temperature.

Página 19 - Hazardous

HMP260 SERIESAppendix 6: Calculation formulas HMP260-U017en-1.52 CORRECTIONS AT HIGH PRESSUREThe saturation pressure of water vapour calculated with e

Página 20 - 14 1997-11-25

BLUGRNYELBLKHMP 265 In automation system galvanically isolated inputsare recommended for current signalsMain board HMP230To control systemSupply volta

Página 21 - 4. COMMISSIONING

HMP263, HMP264 AND HMP265INSTALLATIONToolQtypcsDrawnReviewApprDesignReplacesScaleidTitleArchnoSerial Sheetdoc noCooperator`sDwg noRevLtr QtyChange Rea

Página 22 - OPENED COVER OF THE HMP260

HMP260 SERIESHMP260-U017en-1.5 Appendix 9: ApprovalsAPPROVALS AND SPECIFICATIONS OF THE PROTECTION UNITThe transmitter conforms to the European Standa

Página 23

HMP260 SERIESOperating Manual HMP260-U017en-1.56 1997-11-25The transmitters should be mounted with the sensor head horizontally; thisway, any water co

Página 24

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 7NOTEWhen using the support bar the lead-through piece andthe space between the duct wall an

Página 25 - RX GND T

HMP260 SERIESOperating Manual HMP260-U017en-1.58 1997-11-253. Insert the body of the sensor head into the fitting body and screw thenut manually to th

Página 26 - 20 1997-11-25

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 9Fig. 3.4 Some examples on the installation of the HMP264 sensor head(boring)threaded sleeve

Página 27 - 5. COMMANDS

HMP260 SERIESOperating Manual HMP260-U017en-1.510 1997-11-253.2.3. HMP265 transmitterto be sealedto be insulatedSAFE AREAHAZARDOUS AREAFig. 3.5 Instal

Página 28

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 11a plugged holefor an Exi-approvedreference meterHAZARDOUS AREASAFE AREAFig. 3.7 Mounting t

Página 29 - 1997-11-25 23

HMP260 SERIESOperating Manual HMP260-U017en-1.512 1997-11-25• push back the braid and press it between the two brass disks toachieve a full 360° groun

Página 30

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 13NOTEWhen the cable is grounded as above, the metallic partsof the sensor head, the shield

Página 31

This page intentionally left blank.Back of cover page.

Página 32

HMP260 SERIESOperating Manual HMP260-U017en-1.514 1997-11-25VmA+-CURRENT/VOLTAGEOUTPUTSEXTERNALPOWER SUPPCH1+CH1-CH2+CH2-24 V +OPENED COVER OF THE HM

Página 33

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 154. COMMISSIONINGWhen the HMP260 transmitters leave the factory, their measurement rangesan

Página 34

HMP260 SERIESOperating Manual HMP260-U017en-1.516 1997-11-25CH2 CURRENT OUTPUTCH1 0 ... 1 VOLTAGE OUTPUTCH1CH2CH2CH1X15CURRENT OUTPUTS0 ... 20 / 4 ...

Página 35

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 17Fig. 4.3 Spare jumpers4.3 Connecting the RS 232C serial busFig. 4.4 Serial bus connections

Página 36

HMP260 SERIESOperating Manual HMP260-U017en-1.518 1997-11-25Fig. 4.5 Connection of cablesWhen the serial bus has been connected between the PC and the

Página 37

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 19The RS 232C screw terminal cannot be used if an RS 485/422 serial moduleor a digital curre

Página 38

HMP260 SERIESOperating Manual HMP260-U017en-1.520 1997-11-25jumper must be removed before the transmitter is reset. If the jumper is inplace when powe

Página 39

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 215. COMMANDSAs the HMP260 transmitters are microprocessor based devices, theirconfiguration

Página 40

HMP260 SERIESOperating Manual HMP260-U017en-1.522 1997-11-255.2 LED commandsNOTEIf the transmitter has a display/keypad cover, the LEDcommands cannot

Página 41

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 235.3 Display/keypad commands5.3.1 Display modeIn the display mode the transmitters output m

Página 42

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 iContents1. PRODUCT DESCRIPTION ...

Página 43

HMP260 SERIESOperating Manual HMP260-U017en-1.524 1997-11-25numbers is ended with selecting a blank ' ' and pressing ENT. The last charac-te

Página 44 - 6. CALIBRATION

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 25NOTEThe analogue output jumpers must be set to right places(see Fig. 4.2).5.3.4.2 Selectin

Página 45 - RH 1 point cal. Change the

HMP260 SERIESOperating Manual HMP260-U017en-1.526 1997-11-255.3.5 Output via the serial bus5.3.5.1 Turning the serial interface echo ON/OFF• Select M

Página 46

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 27Selecting stop bits:Full duplex/half duplex:The processor does not allow the following com

Página 47

HMP260 SERIESOperating Manual HMP260-U017en-1.528 1997-11-255.3.5.4 Selecting the output units• Select Unit in the main menu:• Use the arrow keys to s

Página 48

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 295.3.6.1 Setting the serial interface operation mode• Select Mode in the main menu; the fol

Página 49

HMP260 SERIESOperating Manual HMP260-U017en-1.530 1997-11-25• Select More in the main menu, select More in the More menu andselect Mtim in the second

Página 50

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 315.3.7.3 Setting the date• Select More in the main menu; select Date in the More menu:• If

Página 51

HMP260 SERIESOperating Manual HMP260-U017en-1.532 1997-11-255.4 Serial commandsMore detailed descriptions of the serial commands can be found in Appen

Página 52

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 335.4.1.2 Selecting and scaling the analogue output quantitiesDisconnect the security lock j

Página 53

HMP260 SERIESOperating Manual HMP260-U017en-1.5ii 1997-11-255.4. Serial commands ...

Página 54

HMP260 SERIESOperating Manual HMP260-U017en-1.534 1997-11-255.4.2 Output via the serial bus5.4.2.1 Starting the measurement outputR <cr>Starts o

Página 55

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 355.4.2.4 Setting the output interval for the RUN modeINTV xxx yyy <cr>xxx = output

Página 56 - 7. MAINTENANCE

HMP260 SERIESOperating Manual HMP260-U017en-1.536 1997-11-255.4.2.6 Selecting the output unitsUNIT x <cr>x = m(etric units)n(on-metric units)me

Página 57

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 37In POLL mode: measurements output only with command SEND. When inPOLL mode, the output sta

Página 58 - 52 1997-11-25

HMP260 SERIESOperating Manual HMP260-U017en-1.538 1997-11-256. CALIBRATIONThe HMP260 transmitters have been fully calibrated at the factory and theres

Página 59 - 8. TECHNICAL DATA

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 39NOTEIf the sensor has been changed, the calibration has to bedone according to the instruc

Página 60

HMP260 SERIESOperating Manual HMP260-U017en-1.540 1997-11-256.1.1.3 Using LED commandsDisconnect the security lock jumper!• Make sure that the sensors

Página 61

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 41• If you want to see how the sensor stabilizes to the humidity in thecalibrator, enter c &

Página 62 - 56 1997-11-25

HMP260 SERIESOperating Manual HMP260-U017en-1.542 1997-11-25• Wait for 10 minutes.• Connect an ammeter/voltmeter to the analogue outputs (connectorX2)

Página 63 - 1997-11-25 57

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 43• Press any key and enter the second point value and press <cr>.RH : xx.x Ref2 ?

Página 64 - 58 1997-11-25

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 iii8. TECHNICAL DATA...

Página 65 - GUARANTEE

HMP260 SERIESOperating Manual HMP260-U017en-1.544 1997-11-25• Check that the reading corresponds within the desired accuracy to thereading in the cali

Página 66

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 456.2.1 One point offset correction6.2.1.1 Using serial commandsDisconnect the security lock

Página 67

HMP260 SERIESOperating Manual HMP260-U017en-1.546 1997-11-256.2.1.3 Using LED commandsDisconnect the security lock jumper!• Leave the reference instru

Página 68 - 1 ANALOGUE OUTPUT COMMANDS

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 47• The stabilization of the sensor can be monitored by entering c <cr>instead of the

Página 69

HMP260 SERIESOperating Manual HMP260-U017en-1.548 1997-11-256.3 Calibration of analogue outputsThe analogue outputs have been calibrated at the factor

Página 70

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 49• Connect the meter to the output of channel 2 and enter the measuredlower end current/vol

Página 71 - 2 CALIBRATION COMMANDS

HMP260 SERIESOperating Manual HMP260-U017en-1.550 1997-11-257. MAINTENANCE7.1 Reference measurementsReference measurements are necessary for the verif

Página 72

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 51When errors have occurred, the transmitter outputs the error code (see Appen-dix 3 for all

Página 73

HMP260 SERIESOperating Manual HMP260-U017en-1.552 1997-11-25CH1+CH1-TEST POINTSCH1+CH1-mARLX2Fig. 7.4 Circuit diagram of the analogue output current t

Página 74

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 538. TECHNICAL DATASpecifications of the protection unit, see Appendix 9.8.1 Relative humidi

Página 75 - 3 OUTPUT VIA THE SERIAL BUS

HMP260 SERIESOperating Manual HMP260-U017en-1.5iv 1997-11-25This page intentionally left blank.

Página 76

HMP260 SERIESOperating Manual HMP260-U017en-1.554 1997-11-25Accuracies of the calculated quantities depend on the accuracies of the meas-ured quantiti

Página 77

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 55Accuracy of wet bulb temperature (°C)Relative humidityTemp. 102030405060708090100-40 0.20

Página 78

HMP260 SERIESOperating Manual HMP260-U017en-1.556 1997-11-25Connections screw terminals, 0.5 mm2 wires (AWG20), stranded wires recommendedOperating vo

Página 79 - "xxx...xxx"

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 57Ø 13.5Ø 13.5cable length 2000,5000, 10 000 or 15 000 m2970Fig. 8.1 HMP263 sensor head dime

Página 80 - 4800 E 7 1 FDX

HMP260 SERIESOperating Manual HMP260-U017en-1.558 1997-11-258.8 Electromagnetic compatibilityThe emission and immunity tests have been performed accor

Página 81

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 599. OPTIONSMounting flanges for HMP265 aluminiumstainless steel for harsh conditionsInstall

Página 82

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsSERIAL COMMANDS1 ANALOGUE OUTPUT COMMANDS ...

Página 83

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5

Página 84 - 4 OTHERS

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commands1 ANALOGUE OUTPUT COMMANDSAMODE Setting the analogue outputsDisconnect the security lock jum

Página 85

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5ASEL RH T <cr>Ch1 (RH) lo 0.000 %RH ? <cr>Ch1 (RH) hi 100.000 %RH ? <cr

Página 86

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 11. PRODUCT DESCRIPTIONNOTEBefore installing and using the HMP260 transmitterstudy carefully

Página 87

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsThe output scales can also be given directly with the ASCL command.ASCL aaa.a bbb.b ccc.c d

Página 88 - SAFETY SUMMARY

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.52 CALIBRATION COMMANDSCRH Relative humidity calibrationDisconnect the security lock jumper!C

Página 89

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsFCRH Relative humidity calibration after sensor changeDisconnect the security lock jumper!FC

Página 90

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5In one-point offset correction, the Ref2 prompt is acknowledged with <cr>:CT <cr>

Página 91

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsLI Entering linear correction coefficientsDisconnect the security lock jumper!LI <cr>

Página 92

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.53 OUTPUT VIA THE SERIAL BUSR Starting the measurement outputR <cr>Starts output of mea

Página 93

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsThe output mode can be changed with command FORM.DSEND Outputting readings of all connected

Página 94

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5ECHO Turning the serial interface echo ON/OFFECHO xxx <cr>xxx = ON or OFFWhen the ech

Página 95

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsFORM Setting the output formatFORM <cr>"xxx...xxx"? zzz...zzz <cr>xx

Página 96

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5An example of a format suitable for use in Microsoft Excel spreadsheets:FORM <cr>"

Página 97

HMP260 SERIESOperating Manual HMP260-U017en-1.52 1997-11-25OptionsCalculation variables dewpoint temperature, mixing ratio, absolute hu-midity, wet bu

Página 98

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsFDATE Adding date to outputFDATE xxx <cr>xxx = ON or OFFWhen FDATE is activated, the

Página 99

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5The settings can be changed one parameter at a time or all parameters at once:SERI O H <c

Página 100 - 5 SPECIFICATIONS

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsADDR Setting the transmitter addressADDR aa <cr>aa = address (0...99)The address is u

Página 101

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5OPEN aa <cr>SMODE xxxx<cr>aa = address of the transmitterxxxx = STOP, RUN or P

Página 102 - 1 INSTALLATION

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commands4 OTHERSITEST Testing the analogue outputsITEST <cr>orITEST a b <cr>a = current

Página 103 - 2 OPERATION

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5MTIM <cr>Mtim : 4 ? 5 <cr>orMTIM 5 <cr>Mtim : 5PRES Setting the pressure

Página 104

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsor pressure is set to zero using XPRES. After this the pressure stored using commandPRES is

Página 105 - 3 NETWORK CONFIGURATION

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5VERS Name and version of the programmeVERS <cr>For example:VERS <cr>HMPS 260 / x

Página 106 - RX GND TX

HMP260 SERIESHMP260-U017en-1.5 Appendix 2: Safety summarySAFETY SUMMARYIn this summary the following symbols are used:WARNINGA WARNING denotes a hazar

Página 107

HMP260 SERIESAppendix 2: Safety summary HMP260-U017en-1.5CAUTIONOnly fuses with the required current rating and of thespecified type may be used. Do n

Página 108

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 32. TO BE NOTED WHEN MEASURING HUMIDITYIn the measurement of humidity and especially in cali

Página 109 - RH= 24.4 %RH T= 29.1 'C

HMP260 SERIESHMP260-U017en-1.5 Appendix 3: RS 485/422 serial port moduleINSTALLING AND USING THE RS 485/422 SERIAL PORT MODULE1 INSTALLATION...

Página 110

HMP260 SERIESAppendix 3: RS 485/422 serial port module HMP260-U017en-1.51 INSTALLATIONSwitch the transmitter off.Resistors R2, R3 and R4 between conne

Página 111 - ERROR MESSAGES

HMP260 SERIESHMP260-U017en-1.5 Appendix 3: RS 485/422 serial port module2 OPERATIONThe HMP260 transmitters can either be given an address or operated

Página 112

HMP260 SERIESAppendix 3: RS 485/422 serial port module HMP260-U017en-1.5The RS 485/422 module has separate lines for transmitting and receiving, butth

Página 113

HMP260 SERIESHMP260-U017en-1.5 Appendix 3: RS 485/422 serial port module3 NETWORK CONFIGURATIONSingle loop operationBidirectional data on one pair is

Página 114 - 1 CALCULATION

HMP260 SERIESAppendix 3: RS 485/422 serial port module HMP260-U017en-1.5• Set the address of the transmitter; it can be any number between 1and 99. In

Página 115

HMP260 SERIESHMP260-U017en-1.5 Appendix 3: RS 485/422 serial port module• Open the transmitter cover.• Pull out the RS 485/422 serial port module, if

Página 116

HMP260 SERIESAppendix 3: RS 485/422 serial port module HMP260-U017en-1.5NOTE 2The transmitter outputs no prompt (>) after the SMODEPOLL command and

Página 117 - 94-07-04 KKe

HMP260 SERIESHMP260-U017en-1.5 Appendix 3: RS 485/422 serial port module4 CHECKING THE SERIAL PORT NETWORK OPERATIONNormally measurement readings are

Página 118

HMP260 SERIESAppendix 3: RS 485/422 serial port module HMP260-U017en-1.5POLL modeIf a transmitter has been set to POLL mode, it will respond only to c