HMP260 SERIES
HMP260-U017en-1.5 Operating Manual
1997-11-25 27
Selecting stop bits:
Full duplex/half duplex:
The processor does not allow the following combinations:
no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP260 programme will change the number of stop bits to 2
even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the programme changes the number of stop bits to 1
NOTE
The serial bus settings become effective only after reset.
5.3.5.3 Setting the transmitter address
Address is used when more than one transmitters are connected to one serial
bus; this way, it is possible to communicate with one transmitter at a time.
Select Addr in the main menu; the following is displayed:
Pressing ENT returns the programme to the main menu.
Pressing CL deletes the old address; enter the new address with the
arrow keys.
Vista de página 32