Vaisala HMP260 SERIES Manual de Serviço Página 95

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 118
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 94
HMP260 SERIES
Appendix 3: RS 485/422 serial port module HMP260-U017en-1.5
Set the address of the transmitter; it can be any number between 1
and 99. In this example the address is 22:
>addr 22
Address : 22
Set the serial bus settings according to your network specifications.
This setting will become valid after next RESET or power off:
>seri 2400 e 7 1 h
2400 E 7 1 HDX
Set the transmitter in POLL mode:
>smode poll
Serial mode : POLL
NOTE 1
The SMODE command must be given last.
NOTE 2
The transmitter outputs no prompt (>) after the SMODE
POLL command and it only reacts to commands which
include its address.
Check that the transmitter responds to its address:
>send 22
RH= 24.4 %RH T= 29.1 'C
Disconnect the terminal.
Check that the jumpers in connector X4 are in the right places.
Remount the RS 485/422 serial module.
Close the cover.
When all transmitters on the network have been configured, switch
them off.
Dual loop operation
The following procedure must be repeated with all transmitters.
X4
X4
X1 X2
Vista de página 94