Vaisala HMP260 SERIES Manual de Serviço Página 102

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 118
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 101
HMP260 SERIES
Appendix 4: Digital current loop module HMP260-U017en-1.5
1 INSTALLATION
Switch the transmitter off.
Resistors R2, R3 and R4 between connectors X4 and X5 in the component board in
the cover of the transmitter are removed with side-cutting pliers. The module is plugged
in connectors X4 and X5 on the main board of the HMP 260 transmitter; connector X1
on the module board to connector X4 and connector X2 to connector X5.
Connect the data wires to screw terminal X6 on the main board. Switch the power on.
RX GND TX
R2
R3
R4
X6
RX +
TX +
X4
X5
Current loop
module
X1X2
DATA LO OPS
TX -
RX -
X17
Vista de página 101
1 2 ... 97 98 99 100 101 102 103 104 105 106 107 ... 117 118

Comentários a estes Manuais

Sem comentários