20 1997-11-25
jumper must be removed before the transmitter is reset. If the jumper is in
place when power is turned on, the transmitter does not work.
After jumper insertion the transmitter is in STOP mode, ready to receive
commands.
The same method is used when the transmitter is in POLL mode and the user
has forgotten its address.
NOTE
Inserting a jumper in any other place in connector X16
voids the guarantee of the transmitter.