Vaisala HUMICAP HMW90 SERIES Especificações Página 61

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 74
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 60
555
HUMIDITY
HUMIDITY
9
WE MAKE IT EASY.
kele.com 888-397-5353 USA
March 2014
INFRARED DEW POINT & GRAINS TRANSMITTER
9002
Accuracy Dew point: ±3.6°F (2.0°C) @ 77°F
Supply Voltage 24 VAC, ±25%, 2.75 VA
Signal Output simultaneous outputs:
0-10V @ 1000 min
4-20 mA @ 500 max
Relay Output SPDT, 2A @ 24 VAC
Measurement Range Dew point: 27° to 80°F (-3° to 27°C)
Sensing Technology Nondispersive infrared
Calibration In field with nitrogen gas
Adjustments Setpoint: Adjustable, full range
Differential: Adjustable, full range
Altitude Limits Correctable to 5000' (1524m)
Keypads Four-key menu control
Life Expectancy 15 years
Display Two-line, 1/2" (1.27 cm) LCD
Output Range Adjust Within 0° to 80°F (-17 to 26°C)
Operating Humidity 0% to 98%
Operating Temperature 32° to 120°F (0° to 50°C)
Warranty 18 months
DESCRIPTION
The Amphenol Advanced Sensors VaporStat Model
9002 is an absolute humidity transmitter that measures the
amount of water vapor in the air with adversity to temperature
variations. The moisture level is then calculated as dew point
and transmitted simultaneously as 0-10 VDC and 4-20 mA.
A display and keyboard are incorporated for local indication
and parameter adjustment. A contact closure is also
incorporated with keyboard entries for set point and
differential. The Model 9002 is wall-mounted and available
with accessories for duct and OSA applications.
APPLICATION
The Model 9002 sampling chamber measures the infrared
light absorbed by the moisture in the air and processes
it as dew point or grains of moisture per pound of dry air.
The 0-10 VDC or 4-20 mA output ranges are selected at
the keyboard and displayed on the LCD readout. A control
contact energizes at the set point entered on the keyboard.
The Model 9002 is ideally suited for local alarm and signaling
for clean room controls, desiccant system controls, paper mill
controls, and textile process controls as well as condensation
prevention.
SPECIFICATIONS
9002
9002 and
T1508 Enclosure
FEATURES
Dew point accuracy ±3.6°F (±2°C)
0-10V and 4-20 mA output
Digital display or blank cover
Patented infrared absorption platform
Gold-plated element for durability
Built-in keyboard for range adjust
Elevation correction via keyboard
Tamper-resistant keyboard cover
10-minute field calibration with gas