CHAPTER 7__________________________________________________ADJUSTMENT AND CALIBRATION
VAISALA _________________________________________________________________________ 63
correction ? -0.08 <cr>
P1 3. reading ? 698.71 <cr>
correction ? -0.01 <cr>
P1 4. reading ? 801.12 <cr>
correction ? -0.01 <cr>
P1 5. reading ? 900.61 <cr>
correction ? -0.03 <cr>
P1 6. reading ? 947.42 <cr>
correction ? -0.02 <cr>
P1 7. reading ? 1000.10 <cr>
correction ? -0.04 <cr>
P1 8. reading ? 1099.58 <cr>
correction ? -0.04 <cr>
>mpc on <cr>
Multipoint adj: ON
Reading Correction
499.720 -0.070
599.200 -0.080
698.710 -0.010
801.120 -0.010
900.610 -0.030
947.420 -0.020
1000.100 -0.040
1099.580 -0.040
>
NOTE
Always remember to return the switch SW4 to write DISABLE
position (OFF).
Calibration
Calibration is considered not to involve any adjustments. During
calibration, the accuracy of the barometer is verified using a suitable
pressure standard and due corrections are then given in the calibration
certificate together with a description of the international traceability
chain.
After having completed the calibration, the user should store the new
valid date of calibration in the memory of the barometer. This can be
done with the CALD command.