USER'S GUIDE_______________________________________________________________________
44 ___________________________________________________________________ M210194EN-A
T
ABLE 6-1 Commands used in RUN and STOP modes
Function Command
starting output R
stopping output S
single message output SEND
verification message V
self-diagnostics DNOS
pressure minimum/maximum tracking TRACK
pressure limit listing PLIM
resetting the barometer RESET
R Starting the continuous output
R <cr>
This command is used to start continuous outputting in the STOP and
RUN modes and in interval output mode (see page 27).
Example:
>reset <cr>
PTB220 / 2.02
>r <cr>
1013.25 hPa
1013.25 hPa
1013.25 hPa
...
S Stopping the output
S <cr>
This command is used to stop the continuous outputting (activated
with power-up in RUN mode or using the RUN command).
Example:
>reset <cr>
PTB220 / 2.02
>r <cr>
1013.25 hPa
1013.25 hPa
1013.25 hPa
s <cr> (text invisible)
>