USER'S GUIDE_______________________________________________________________________
34 ___________________________________________________________________ M210194EN-A
An external power control can be used to switch the PTB220
barometer ON/OFF. The control is activated with the switch SW2 in
position ON as shown in FIGURE 4-4. The TTL signals are: 0VDC OFF
and 5VDC ON.
ON
OFF
SW1
SW2
SW3
SW4
SW5
SW6
SW7
SW8
S
FIGURE 4-4 External power control ON
The pulse output mode is activated and the barometer starts to wait for
an external trigger pulse by turning the switch SW3 in position ON as
in
FIGURE 4-5. See Appendix D for quick reference information on how
to set and use the pulse output mode of the PTB220 series digital
barometers.
ON
OFF
SW1
SW2
SW3
SW4
SW5
SW6
SW7
SW8
S
FIGURE 4-5 Pulse mode ON
NOTE
Switches SW6 and SW7 are usually in position OFF. However, for
some specific uses they may be turned ON to emulate other
barometers (see page 103 for further details).