USER'S GUIDE____________________________________________________________________
70 __________________________________________________________________ M210762EN-D
The configuration of the Display Alarms can only be done using the
display/keypad option.
An alarm is activated when the selected quantity goes higher than the
high limit, or lower than the low limit, much in the same way as the
relays. When an alarm is activated, an alarm note is displayed on the
display, and the lights of the display will blink.
0802-041
Figure 44 Display Alarm Active
Multiple alarms can be active at the same time; the alarm that was
triggered first will be shown on the display. The next active alarm is
revealed when the currently shown alarm is acknowledged by pressing
the OK button.
Note that activated alarms are only shown on the screen. There are no
alarm messages output to the serial line, or markers placed in the graph
data. After an alarm has been acknowledged, you must refer to the data
graphs to see when the measured quantities have exceeded the limits.
Configuring a Display Alarm
1. Enter the MAIN MENU by pressing an arrow key on the keypad.
2. Use the arrow keys to select Display, followed by Alarms, to open
the Display Alarms menu. The Display Alarms menu shows the
currently enabled and disabled alarms.
0802-069
Figure 45 Display Alarms
3. Use the arrow keys to select an alarm to configure. The alarm
editing page opens.