Figure 66 Error Indicator and Error Message
Number refers to Figure 66 above:
1 = Error Indicator
- The error indicator disappears when the error state is over and you
have checked the error message. Press the INFO button to display the
error message.
Table 30 Error Messages
Error Code Error Message Action
0 Humidity sensor
measurement malfunction.
Check the integrity of the humidity probe and the
probe cable. Clean the probe from dirt, water,
ice or other contaminants.
1 Humidity sensor short circuit Check the integrity of the humidity probe and the
probe cable. Clean the probe from dirt, water,
ice or other contaminants.
2 Humidity sensor open circuit Check the integrity of the humidity probe and the
probe cable.
3 Temperature sensor open
circuit.
Check the integrity of the humidity probe and the
probe cable.
4 Temperature sensor short
circuit.
Check the integrity of the humidity probe and the
probe cable. Clean the probe from dirt water, ice
or other contaminants.
5 Temperature measurement
malfunction
Check the integrity of the humidity probe and the
probe cable. Clean the probe from dirt water, ice
or other contaminants.
6 Temperature sensor current
leak.
Check the integrity of the humidity probe and the
probe cables. Clean the probes from dirt, water,
ice or other contaminants.
7 Internal ADC read error Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.
9 Checksum error in the
internal configuration
memory
Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.
10 Internal EEPROM read
error
Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.
11 Internal EEPROM write
error
Internal transmitter failure. Remove the
transmitter and return the faulty unit to Vaisala
Service.
12 … 13 Add-on module 1 (or 2)
connection failure
Turn off the power and check the module
connection. Turn on the power.