Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 115
0610-077
Figure 61 FAULT/ONLINE STATUS Relay Output Modes
FAULT/ONLINE STATUS relays are usually used in conjunction with
an analog output to obtain validity information for the output value.
NOTE
If transmitter loses its power, all status-based relays are released
similarly to the case of an instrument failure.
Enabling/Disabling Relays
You can deactivate the relay outputs for example for service purposes of
your system.
Indication Led Operation
Relay is activated: LED is lit
Relay is not activated: LED is not lit