Vaisala PTU300 Guia do Utilizador Página 130

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 217
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 129
User's Guide _______________________________________________________________________
128 __________________________________________________________________ M210796EN-G
Example: 0 ... 5 V voltage output selected for channel 1 and 4...20 mA
selected for channel 2.
OFF
ON
Selection
1
2
Voltage output selected
3
4
5
6
0 ... 5 V selected
7
8
1
Current output selected
2
3
4
4 ... 20 mA selected
5
6
7
8
NOTE
If you have customized the error output setting (
AERR), check that the
set error values are still valid after changing the output
mode/range. See
section
Analog Output Fault Indication Setting on page 131.
Analog Output Quantities
NOTE
For best accuracy, pressure must always be output using Ch3, if
available.
Use the display/keypad to change and scale the analog output quantities.
1. Press any of the arrow buttons to open the Main Menu.
2. Select Interfaces and press the right arrow button.
3. Select Analog outputs and press the right arrow button.
4. Select Output 1/2/3 and press the right arrow button.
5. Select Quantity and press the CHANGE key.
6. Select the quantity by using the up/down arrow buttons. Press the
SELECT key to confirm your selection.
7. Select Scale, lower limit, by pressing the up/down arrow buttons.
Press the SET key. Adjust the lower limit value by pressing the
arrow buttons up/down/left/right. Press the OK key to confirm your
setting.
Vista de página 129
1 2 ... 125 126 127 128 129 130 131 132 133 134 135 ... 216 217

Comentários a estes Manuais

Sem comentários