11.3.3 General
Operating temperature range -20...+60 °C
sensor head -40...+100 °C (temporarily + 180 °C)
Storage temperature range -40...+80 °C
Material:
handle ABS plastic
pipe stainless steel
cable PVC spiral cable
Housing classification (electronics) IP65 (NEMA 4)
Weight 450 g (incl. package)
11.4 Accuracy of the calculated quantities
Dewpoint temperature, mixing ratio, absolute humidity and wet bulb tempera-
ture are calculated from the measured relative humidity and temperature
values. The accuracy of the calculated quantities depends on the calibration of
the probe and on performing the measurement correctly. In tables below the
accuracies of the measured values are ±2 %RH and ±0.2°C.
11.4.1 Accuracies in metric units
Accuracy of dewpoint temperature (°C)
RH/%
10 20 30 40 50 60 70 80 90 100
T/°C
-40
1.82 1.00 0.74 0.61 0.53 0.48 0.44 0.42 - -
-20
2.09 1.14 0.83 0.68 0.59 0.53 0.49 0.45 - -
0
2.51 1.37 1.00 0.81 0.70 0.63 0.57 0.53 0.50 0.48
20
2.87 1.56 1.13 0.92 0.79 0.70 0.64 0.59 0.55 0.53
40
3.24 1.76 1.27 1.03 0.88 0.78 0.71 0.65 0.61 0.58
60
3.60 1.96 1.42 1.14 0.97 0.86 0.78 0.72 0.67 0.64
80
4.01 2.18 1.58 1.27 1.08 0.95 0.86 0.79 0.74 0.70
100
4.42 2.41 1.74 1.40 1.19 1.05 0.95 0.87 0.81 0.76
120
4.86 2.66 1.92 1.54 1.31 1.16 1.04 0.96 0.89 0.84
140
5.31 2.91 2.10 1.69 1.44 1.27 1.14 1.05 0.97 0.91
160
5.80 3.18 2.30 1.85 1.57 1.38 1.24 1.14 1.06 0.99