USER'S GUIDE____________________________________________________________________
132 _________________________________________________________________ M210796EN-E
- Select Automatic purge and turn it on or off by pressing the
ON/OFF key.
- Select Interval and press the SET key. Set the purge interval
and the unit (hour/day) by using the arrow buttons. The interval
must be set between 1 hour...10 days. Press the OK key.
- Select Start-up purge and press the ON/OFF key.
- Start manual purge by selecting Manual purge and pressing the
START key.
4. Press the EXIT key to return to the basic display.
0706-004
Figure 65 Performing Chemical Purge
Using Serial Line
PURGE
Enter the PURGE command to start chemical purge immediately.
>purge
Purge started, press any key to abort.
>
The prompt '>' appears when the heating period is over. However, the
transmitter outputs are locked to the values measured before performing
chemical purge until the settling time is over.
With PUR command you can enable or disable automatic and power-up
chemical purge and set the interval for automatic purge. If the sensor is
exposed to chemicals it is recommended to have the chemical purge done
at least once in 720 min (=12 hours). In applications where the chemical
exposure is not likely, the interval can be longer.
It is not recommended to change duration, settling, temperature or temp.
difference.
PUR
Type PUR and press ENTER
to proceed. The maximum interval is
14400 minutes (=10 days).