Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 95
Table 25 Other Commands
Command Description
? Output information about the device
?? Output information about the device in
POLL mode
ECHO [ON/OFF] Turn the serial interface echo ON/OFF
ERRS List present transmitter errors
FILT Set the result filtering
FIND All devices in POLL mode send their
addresses
HELP List the currently available commands
LIGHT Set the display backlight mode
LOCK Lock the menu or disable the keypad
MODBUS View MODBUS diagnostic counters
PRES [hPa] Set the value for pressure compensations
VERS Display the software version information
XHEAT Sensor heating
XPRES [hPa] Set the value for pressure compensations,
temporarily
Getting Measurement Message from Serial Line
Starting Continuous Outputting
Use the R command to start the continuous output of measurements.
R<cr>
Example:
>r
RH= 60.5 %RH T= 23.7 'C Tdf= 15.6 'C Td= 15.6 'C a= 13.0
g/m3 x= 11.1 g/kg Tw= 18.5 'C H2O= 17889 ppmV pw=
17.81 hPa pws= 29.43 hPa h= 52.3 kJ/kg dT= 8.1 'C
If a value is too long to fit to the allocated space in the output, or if there
is an error in outputting the quantity, the value is displayed with stars ‘*’.
Example:
>r
RH=***.* %RH T= 31.0 'C
You can change the format of the output with the following commands:
- Outputting interval can be changed with the INTV command.
- Output message format can be changed with the FORM command.