8. Select the upper limit by pressing the ▲▼arrow buttons. Use the
arrow buttons to set the upper limit value. Press SET to confirm
your selection. Press OK to confirm your setting.
9. Press EXIT to return to the basic display.
AMODE/ASEL
Use the serial line to select and scale the analog output quantities. Check
the analog output modes with the AMODE command.
AMODE<cr>
Example:
>amode
Ch1 output : 0...1V
Ch2 output : 0...1V
>
Select and scale the quantities for the analog outputs with the command
ASEL. Note that the optional quantities can be selected only if they have
been selected when ordering the device.
ASEL [xxx yyy zzz]<cr>
where
xxx = Quantity of channel 1
yyy = Quantity of channel 2
zzz = Quantity of the optional analog output channel 3
Always enter the quantities for all outputs. For quantities and their
abbreviations see Table 3 on page 19 and Table 4 on page 19.
Use the ASEL [xxx yyy] command as shown in the exam
ple below when
using a device with two analog outputs.
Example:
>asel rh t
Ch1 (RH ) low : 0.00 %RH ? 0
Ch1 (RH ) high : 100.00 %RH ? 100
Ch2 (T ) low : -40.00 'C ? -50
Ch2 (T ) high : 60.00 'C ? 80
>