Vaisala HUMICAP DMT132 Guia do Utilizador Página 50

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 49
USER'S GUIDE____________________________________________________________________
48 __________________________________________________________________ M211289EN-A
Restore Factory Settings
Use the FRESTORE command to restore the factory settings to the
transmitter. All user settings will be lost, and the transmitter will return to
the factory calibrated state (user adjustment is cleared).
FRESTORE<cr>
Example:
>frestore
Restoring factory defaults...done
Saving settings...done
>
CAUTION
If you have changed the HUMICAP® sensor yourself, be aware that the
FRESTORE command clears all of the user adjustment parameters,
including the capacitance offset and gain that are set by the
CRH
command.
If you want to avoid having to perform the capacitance adjustment
procedure again, record the adjustment parameters shown by the
L
command before performing the
FRESTORE command. You can
afterwards restore the adjustment parameters using the
LI command.
See page 46.
NOTE
If the HUMICAP® sensor has been changed at a Vaisala Service Center
the factory calibration has been updated to match the new sensor, and it
will not be affected by the
FRESTORE command.
Vista de página 49
1 2 ... 45 46 47 48 49 50 51 52 53 54 55 ... 73 74

Comentários a estes Manuais

Sem comentários