Vaisala HUMICAP DMT132 Guia do Utilizador Página 13

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 12
Chapter 2 __________________________________________________________ Product Overview
VAISALA _______________________________________________________________________ 11
Transmitter Structure
The structure of the DMT132 is shown in Figure 2 on page 11. The
transmitter body does not have user serviceable parts inside, and is not
designed to be opened. Opening the transmitter will void the warranty.
1101-083
Figure 2 DMT132 Transmitter Structure
where
1 = 30 mm nut
2 = Transmitter body
3 = 4-pin M8 connector I: analog output and operating power
4 = 4-pin M8 connector II (shown with protective cap):
Connection for service port (RS-485, service use only ) and
LED indicator plug
5 = Sealing ring (included with every transmitter)
6 = Connection thread (ISO G1/2")
7 = Tube filter that protects the HUMICAP® and Pt1000 sensors
When the transmitter is delivered, the filter is protected by a yellow
transport protection cap that keeps the sensor dry. The transport
protection cap should be left on the transmitter during storage. Remove
the transport protection cap before installing the transmitter.
Vista de página 12
1 2 ... 8 9 10 11 12 13 14 15 16 17 18 ... 73 74

Comentários a estes Manuais

Sem comentários