User's Guide ______________________________________________________________________
122 _________________________________________________________________ M210205EN-D
Wetted parts' materials Stainless steel (AISI
316L), silicone
elastomer, Vectra LCP,
tantalum, quartz
Housing classification IP31 (NEMA 2)
Mechanical piping connections Swagelok 6 mm/ /¼"
Storage temperature range -40...+70 ºC
Weight 7 kg
Minimum heat removal capacity
of the cooling water system 250 W @ 5 ºC
Options for DMP501
Pump 0...1 l/min,
0... 200 mbar
differential pressure
Heated internal gas sampling
Connection cable to DMI500 2.5 or 10 meters
Temperature probe
Measurement range -40…+80 ºC
(-40…+176 ºF)
Typical accuracy ±0.1 ºC (±0.18 ºF)
Sensor Pt100 PRT DIN
IEC 751 class ¼ B
Cable length 2 m /5 m /10 m
DMI500 User Interface Unit
Operating voltage 100-230 VAC
Operating current 5A-3A
Operating power max 460 W
Display B/W LCD with
backlight
Menu languages English, German,
French, Finnish and
Spanish
Operating temperature 0...+40 ºC
Storage temperature range -40...+70 ºC
Electronics housing material PPE + PS plastic,
aluminium, stainless
steel
Housing classification IP31 (NEMA 2)
Weight 7.5 kg
Comentários a estes Manuais