Chapter 5 _______________________________________________________________ Maintenance
VAISALA ________________________________________________________________________ 47
CHAPTER 5
MAINTENANCE
This chapter provides information that is needed in basic maintenance of
HMT120.
Replacing the HUMICAP
®
Sensor
Fixed and Remote Probe Models
1. Remove the filter (plastic grid or sintered stainless steel).
2. Remove the damaged sensor and insert a new one.
3. Recalibrate the probe.
4. Replace a dirty filter to ensure a maximum lifetime and a fast
response for the sensor. Do not attempt to clean the filter.