Chapter 6 ___________________________________________________ Calibration and Adjustment
VAISALA _______________________________________________________________________ 53
6. When the reading is stable, give the reference humidity after the
question mark and press ENTER.
crh
RH : 25.21 Ref1 ?
RH : 25.22 Ref1 ?
RH : 25.22 Ref1 ? 30.1
Press any key when ready ...
RH : 25.19 Ref2 ?
OK
7. OK indicates that the adjustment has succeeded and the new
calibration coefficients are calculated and stored.
8. Take the probe out of the reference conditions and replace the
filter.
Using MI70
To adjust the relative humidity:
1. Connect MI70 to MMT162.
2. Turn MI70 on.
3. On the MI70, go to Menu Functions Adjustments.