USER'S GUIDE____________________________________________________________________
Adjusting Temperature
Temperature Adjustment of DM70
The temperature adjustment can be done using the one-point or two-
point adjustment. The one-point adjustment should be done within the
temperature range in which the device is most often used.
To go to the adjustment mode, do the following:
1. Press the Calibration button of the probe handle (see number 4
in Figure 1 on page 11) using a tool with a thin, sharp point, for
example, a small screwdriver. The indicator will switch into the
adjustment mode. To confirm calibration, press OK.
2. Select T
d/f
and press Select.
3. Select T and press Select.
4. To select the adjustment method, first press Adjust and then
the adjustment method, either 1-point adjustment or 2-point
adjustment.
5. Depending on your selection, continue from section One-Point
Adjustment below or Two-Point Adjustment on page 75.
One-Point Adjustment
1. After selecting 1-point adjustment, press Select to
continue.
2. Set the probe to the reference temperature. You can follow the
stabilization from the graph display by pressing
Graph.
When the reading has stabilized, press Ready.
3. Use the arrow buttons to give the reference temperature value
and then press OK.
4. Confirm the adjustment by pressing
Yes. (If you press
No, the adjustment mode display returns and no changes are
made.) The adjustment has now been completed.
5. To exit the adjustment mode, press
Back.
6. To return to the basic display, press Exit.
74 __________________________________________________________________ M010091EN-F