Chapter 5 _____________________________________________________________ User Interface
3. To change the recording interval, select Interval and press
Set.
4. Select the measurement interval with the arrow buttons. The
measurement intervals and the maximum recording times are
shown in Table 1 on page 33.
Table 1 Measurement Intervals and Maximum
Recording Times
Maximum Recording Time (Memory Full) Measure-ment
Interval
1 quantity 2 quantity 3 quantity
1 s 45 min 22 min 15 min
5 s 3 h 113 min 75 min
15 s 11 h 5 h 3 h
30 s 22 h 11 h 7 h
1 min 45 h 22 h 15 h
5 min 9 days 4 days 3 days
15 min 28 days 14 days 9 days
30 min 56 days 28 days 18 days
1 h 113 days 56 days 37 days
3 h 339 days 169 days 112 days
12 h 1359 days 678 days 451 days
5. Press Select.
6. To set the duration of the recording, select Duration and then
press Set. Select the duration with the arrow buttons and
press Set.
7. To start recording, select Start/Stop recording, press Start.
If you choose Memory full, you will see the maximum recording
time on the display. You can also delete files to empty the
memory. To accept the maximum recording time, press
Start again.
8. You can switch off DM70 during recording to save the battery.
A message will appear on the display to inform that recording
will continue undisturbed even when the power is turned off. If
the indicator is switched off during recording, the progress bar is
shown on the display every 10 seconds (continuously if the
charger is connected). The bar indicates the amount of recorded
data.
CAUTION
Do not disconnect the probe when the data recording is on, even if
the indicator is turned off. This may cause loss of recorded data.
VAISALA _______________________________________________________________________ 33