Chapter 7 ____________________________________________________________ Technical data
VAISALA ______________________________________________________________________ 147
Inputs and Outputs
Operating voltage 10...35 VDC, 24 VAC
with optional power supply module 100...240 VAC, 50/60 Hz
Start-up time after power-up 3 s
Power consumption @ 20 °C (Uin 24VDC)
RS-232 max 28 mA
Uout 3 x 0...1V / 0...5V / 0...10V max 33 mA
Iout 3 x 0...20 mA max 63 mA
display and backlight + 20 mA
during chemical purge + 110 mA max
during probe heating (HMT337) + 120 mA max
Settling time at power-up (one sensor)
class A 4 s
class B 3 s
Analog outputs (2 standard, 3rd optional)
current output 0...20 mA, 4...20 mA
voltage output 0...1 V, 0...5 V, 0...10 V
Humidity and temperature
Accuracy of analog outputs at 20 °C ± 0.05 % full scale
Temperature dependency of the
analog outputs ± 0.005 %/°C full scale
Pressure
500...1100 hPa 50...1100 hPa
Accuracy of analog outputs at 20 °C 0.30 hPa 0.4 hPa
Temperature dependency of the
analog outputs 0.6 hPa 0.75 hPa
External loads
current ouputs RL < 500 ohm
0... 1V output RL > 2 kohm
0... 5V and 0... 10V outputs RL > 10 kohm
Max wire size 0.5 mm2 (AWG 20) stranded
wires recommended
Digital outputs RS-232, RS-485 (optional)
Relay outputs (optional) 0.5 A, 250 VAC, SPDT
Display (optional) LCD with backlight, graphic
trend display
Menu languages English, French, Spanish,
German, Japanese, Swedish,
Finnish
Mechanics
Cable bushing M20x1.5 For cable diameter
8...11mm/0.31..0.43"
Conduit fitting 1/2”NPT
User cable connector (optional) M12 series 8- pin (male)