User's Guide ______________________________________________________________________
136 _________________________________________________________________ M210796EN-A
CT
or for additional T probe:
CTA
4. Type C and press ENTER a few times to check if the reading is
stabilized. Let the reading stabilize, give the reference
temperature after the question mark and press ENTER three
times.
When having another reference temperature (2-point calibration) press
ENTER only twice and insert the probe to the second reference.
When the reading is stabilized, give the second reference temperature
after the question mark and press ENTER. Please, note that the
difference between the two temperature references must be at least
30 ºC.
Example (1-point adjustment):
>ct
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? 16.0
Press any key when ready ...
T : 16.06 Ref2 ?
OK
>
5. OK indicates that the calibration has succeeded.Give the
calibration information (date and text) to the transmitter's
memory, see the serial commands CTEXT and CDATE.
6. Press the ADJ button on the motherboard to close the
adjustment mode.
7. Take the probe out of the reference conditions and replace the
filter.