Vaisala PTB210 SERIES Manual de Instruções Página 110

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 137
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 109
BGC-FTIR Sensors | PTB 210 25
Operation
The PTB210 can be operated through a serial line with any terminal emulator
program and the commands given in the following pages. Note to use the right
parameters for communication with the RS232-Port on your PC or laptop.
(baud rate 9600, parity even, data bits 7, stop bits 1)
General and Communication
The general command format is:
.ZZZ.xxx<cr>
ZZZ = event
xxx = given value of a parameter
If there are several barometers on same line, it is necessary to use an ID-code:
123.ZZZ.xxx<cr>
The default ID-code is 0. If necessary, you can change the ID:
.ID.xxx<cr>
xxx = identification code (max. 15 characters)
To set the baud rate of the barometer use the command:
.BAUD.xxxx<cr>
xxxx = baud rate (1200...19200
NOTE: The barometer does not echo the given commands nor confirm the
new settings by echoing them. However, the settings can be checked
with the command ?<cr>. Make sure to reset the barometer with
the command .RESET<cr> to activate the new settings.
Vista de página 109
1 2 ... 105 106 107 108 109 110 111 112 113 114 115 ... 136 137

Comentários a estes Manuais

Sem comentários