Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 29
Table 9 Calibration and Adjustment Commands
Command Description
# Force connection settings 19200 baud, N, 8, 1
ACAL [0/1] Analog output calibration
CDATE Set the calibration date
CRH Relative humidity calibration
CT Temperature calibration *
CTA Additional temperature probe calibration *
FCRH Relative humidity calibration after sensor
change
CTEXT Give the text to calibration information field
L Displays user adjustment parameters
LI Displays new values for the user adjustment
parameters
Table 10 Setting and Testing the Analog Outputs
Command Description
AMODE View the analog output modes
ASEL Set or view analog output quantities and
scaling
AERR Change the analog error output values
ATEST Force the analog outputs to entered values
Table 11 Other Commands
Command Description
? Output information about the device
?? Output information about the device in POLL
mode *
ERRS List present probe errors
FILT [0.1…1] Set the result filtering
HELP Lists the available commands
PRES [hPa] Set the value for pressure compensations
RESET Resets the probe.
VERS Display the software version information
XHEAT Sensor heating *
XPRES [bar] Set the value for pressure compensations,
temporarily