Analog Output Test
Analog outputs can be tested with dip switch 1, see DIP Switch Functions
on page 45 for details.
Technical Support
For technical questions, contact the Vaisala technical support by e-mail at
information:
- Name and model of the product in question
- Serial number of the product
- Name and location of the installation site
- Name and contact information of a technically competent person who
can provide further information on the problem.
Product Returns
If the product must be returned for service, see
www.vaisala.com/returns.
For contact information of Vaisala Service Centers, see
www.vaisala.com/servicecenters.
Table 7 Error Codes
Error Code
on Display
Error Code
on Serial Line
Typical Cause
RH error RH error HUMICAP® humidity sensor open
circut or short circut. Can be
caused by condensed water, or
damage to sensor or probe.
T error T error Pt100 temperature sensor open
circut or short circuit. Can be
caused by condensed water, or
damage to sensor or probe.
ADC error Prb error Probe module disconnected.
CPU EEPROM error EE error Hardware failure in transmitter.
PRB EEPROM error EEP error Hardware failure in probe module.