Vaisala HMT360Series Guia do Utilizador Página 57

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 56
Chapter 4 ________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 55
SEND Outputting Measurement Values
Syntax: SEND<cr>
This command outputs the measured values in one point.
R Activating Continuous Output
Syntax: R<cr>
With the command R the transmitter outputs measured values
continuously.
S Stopping Continuous Output
Syntax: S<cr>
The continuous outputting is stopped with the command S. If outputting is
active, this command is not echoed.
INTV Setting Output Interval
Syntax: INTV [n xxx]<cr>
Sets the output interval when the transmitter outputs measurement
readings. The time interval is used when the continuous output is active.
Example, the output interval is set to 10 minutes:
>intv 10 min
Output interval: 10 MIN
>
where
n = 1 ... 255
xxx = S, MIN or H
Vista de página 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 96 97

Comentários a estes Manuais

Sem comentários